1 General

The light-curing unit egger eLED.LP3 is suitable for a
number of uses in the acoustician shop, during home
visits and in the otoplastic laboratory. Service operations
can be executed immediately right in the shop – fast,
efficiently, no lengthy preparation needed. So you can
enjoy more independence while offering your customer
the best possible service.

2 Safety Regulations

!
The electrical light-curing unit egger eLED.LP3
is used for the polymerization of UV-curing mo-
deling pastes and lacquers. A use other than described
in the operating instructions may lead to injuries of the
operator or of third persons.
Furthermore, the device or other material assets may be
damaged. The device must be used in technically perfect
condition only and the safety regulations must be strictly
complied with.
Warning hot surfaces – do not touch the LED
reflectors – risk of burns!
Work area identification,
Risk class 3,
according to DIN EN 62471.
At a distance of less than 20 cm from the lamp,
the exposure limits for UV radiation are exceeded.
For the egger eLED.LP3 light polymerization unit, the
operator must prepare a risk assessment and operating
instructions.
On self-accomplished repairs or if other spare parts are
used, all warranty and liability claims will become extinct!
!

2.1 For a Safety Conscious Work Environment

• Protect the unit from dust. Use the unit on dry wor-
king places only.
• Store the unit in closed rooms.
• Qualified and trained persons must use the device only.
Do not look directly into the lamp light!
• Children must be kept away from the unit.
• Surrounding temperature between 10 °C and 30 °C.
Relative air humidity between 10 % and 70 %.
• Before putting the device into operation, the user
must read the operating instructions and ask the
manufacturer in case of any questions.
• Use the unit in perfect and clean condition only and
by observing the operating instructions.
• Connect the device to permissible sockets only.
• Use the included cable only.
• Please adhere to the acknowledged safety regulations.
• This device is used for the processing of UV-curing
materials in a spectral range of 360 – 400 nm.
• Please also adhere to the processing instructions for
each material used.
• Disconnect the unit from electricity supply prior to
any service or maintenance.
• Do not pull the mains plug out of the socket at the
mains cable.
• Should the mains cable be damaged, it must be repla-
ced at once. Defective mains cables must not be used.
• The unit may not be opened by operators. There are
no interchangeable parts such as fuses inside.
• Repairs may only be executed by the manufacturer.
!
Advice for cleaning: Clean the unit with a moist
cloth only – do not use solvents, do not douse the
unit into water.