Troubleshooting

PROBLEM
LCD display
shows
abnormal
result
Low battery symbol Low battery
LCD shows EE
LCD flashing EE
LCD shows EE EP
No response
when you press
button or load
battery.
9

Care and handling

1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and direct sunlight. Do not immerse the monitor in water as this will result in
damage to the monitor.
3. If this monitor is stored near freezing, allow it to acclimatize at room temperature before use.
4. Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the batteries.
6. It is recommended the performance should be checked every 2 years or after repair. Please contact the
service center.
7. If the monitor becomes dirty, please clean it with a soft dry cloth. Do not use any abrasive or volatile cleaners.

Explanation of symbols

Caution - Consult accompanying documents
* The functioning of the unit can be affected by interference from local electromagnetic
disturbance..
Application part type BF
The appliance should be disposed of according to Regulation 2002/96/EC-WEEE (Waste
Electrical and Electronic Equipment).
Manufactured in accordance with international standards. The instrument is in line with the
EU Medical Products Directive MDD 93/42/EC.
POSSIBLE CAUSE
The cuff position is not correct or
cuff is not properly tightened
Body posture was not correct during
testing
Talking, arm or body movement,
irritation, excitement or tension
during testing.
Irregular heartbeat (arrhythmia)
Pneumatic system blocked or cuff is
too tight during inflation
Pneumatic system leakage or cuff is
too loose during inflation
Fail to detect systolic pressure
Fail to detect diastolic pressure
Device parameter checking error
Incorrect operation or strong
electromagnetic interference.
SOLUTION
Apply the cuff correctly and try again
Review the "BODY POSTURE DURING
MEASUREMENT" sections of the
instructions and re-test.
Re-test when calm and without
speaking or moving during the test
It is inappropriate for people with serious
arrhythmia to use this blood pressure
monitor.
Change batteries
Apply the cuff correctly and try again
Don't move and try again.
Take out batteries for five minutes and
then reinstall all batteries. If the monitor
still does not function normally, please
contact the local distributor or the
factory.