Manual dosing (F2)
The display shows F2, press OK and the pump starts pumping.
The tube fills slowly with water and the water will flow out of the other tube.
Press OK to stop pumping.
Press  to move on to the calibration of the pump (F3).
Calibrating the pump (F3)
Before calibrating, make sure there is no air in the tubes and that the measuring cup is empty.
The display shows F3, press OK to start the pump.
The display will show the calibration value incrementing
Press OK to stop the pump when the water level in the measuring cup reaches 100 ml.
Press  to save the calibration value or press  to save and proceed to setting the reservoir content (F4).
NOTICE! Repeat calibration regularly to ensure correct dosing.
Setting the content of the reservoir (F4)
The display shows F4, press OK
Press  or  to set the amount of dosing medium (1-9999 ml) in the reservoir
By pressing OK you can switch between
Press  to save the setting or press  to save and proceed to setting the dosing quantity (F5).
If the liquid in the reservoir is insufficient for the next dosage, the unit emits an alarm signal.
End the alarm signal by pressing OK. Refill the reservoir and set the reservoir content (F4) again.
If the alarm goes off during programming, switch the pump off and on again briefly and then press OK to stop the
alarm.
Setting the dosing quantity (F5)
The display shows F5, press OK
Press  or  to set the required dosing quantity (1-9999 ml)
By pressing OK you can switch between
Press  to save the setting or press  to save and to continue with setting the dosing interval (F6).
Setting the dosing interval (F6)
The display shows F6, press OK
Press  or  to set the time between the doses (between 1 hour and 72 hours)
Press  to save the setting. The pump will immediately dose the set quantity under F5. The next dose will follow
according to the set interval time under F6.
NOTICE! If you do not want the pump to start dosing immediately, press  first and then press .
The dosing interval set under F6 is saved and will not start dosing until after the time set.
The set interval time therefore starts immediately.
Maintenance
• Clean the outside of the dosing pump with a damp cloth.
• The pump head, pump motor and tubes are subject to wear and tear and, depending on their use, must be
replaced after a while. This is normal and is not covered by the warranty.
• Replacement parts are available from your dealer.
NOTICE! NEVER lubricate the pump!