Menu „Config"
•
Language: Select from English or German.
The input is done with the arrow keys left and right.
•
Contrast: Adjusts the contrast of the LCD display.
The setting between 3 and 20 is done with the arrow keys left and right.
•
Time: The current time is displayed. To set the time, press the OK button.
The clock is stopped and the arrow keys left and right change the hours.
By pressing the OK button again you switch to the minutes. The arrow
keys left and right change the minutes now. Pressing the OK button will
change to the seconds. In order to take over the adjusted time, move to
a different configuration entry by means of the arrow keys up or down.
(Please also note the instructions in section Changing the battery,
page 42.)
•
AutoStart1 and AutoStart2: These menu items show whether the auto-
start of a program is active at a configured time. By pressing the OK but-
ton you switch to the screen which provides the appropriate configuration
(see chapter Configuring the Auto-Start function, page 52)
•
Const amps: enables or disables the constant current control.
The input is done with the arrow keys left and right.
If no is selected, the intensity of all the frequency programs (including
those, that are stored on the chip cards) is usually set on the voltage and
does not have a constant current control.
•
Speaker: the tone pitch of the internal audio signal.
Use the arrow keys left and right to choose 4 or 5 kHz.
•
HW Rev.: Displays the hardware version of your device.
All manuals and user guides at all-guides.com
This menu is used to configure the device. The settings
made here are retained even after switching off the device
or changing the battery.
Diamond Shield IE Crystal
Instruction Manual