Troubleshooting

Clean the condensate drain when necessary. If it is blocked, the condensate
collects on the bottom of the refrigerator.
➤ Clean the door seals regularly.
Freestanding version: Clean the grid of the condenser and the compressor on the
back of the refrigerator with a vacuum cleaner regularly.
9
Troubleshooting
Water in or under the refrigerator
Fault
Water is leaking into the
inside of the
refrigerator.
Water is leaking onto
the floor.
Display signals
Signal
General error
The display blinks each
second.
Sensor failure
The display blinks and
shows "E1".
Low battery
The display blinks and
shows "E2" alternately
with the current tem-
perature.
20
Possible cause
The condensate drain duct is
clogged.
Medicines are preventing the water
from flowing to the collector.
There is too much water in the water
container.
Explanation
The display shows alternately the current inner temperature and
the active error code (fig. 2 1, page 3) (E1 or E2, see below).
The minimum and maximum values will be kept as the last known
values.
The minimum and maximum inner temperatures are still visible
and will be updated from the current inner temperature reading.
HC302, HC502
Remedy
Clean the drain.
Make sure that medicines
do not touch the rear
wall.
Clean the water
container.