3.

Unpacking / Assembly

3.1.
Unpacking the CA Basic EVO
Check whether the package contents are complete and that nothing has been damaged during transport.
The complete system, including accessories, contains the following components:
Basic EVO unit with P R SL filter
Exchangeable battery
Comfort padded belt Standard
Comfort padded belt Super
Flow indicator
Battery charger (multi plug)
User manual
3.2.
Assembly
1. Connect the battery to the PAPR unit.
2. Connect the hose to the PAPR unit and make sure the connection is tight enough.
3. Check if the filter is properly attached.
4. Connect the hose to the headtop.
4.

Before use

4.1.
Control procedure before every use - make sure that:
1. All components are in good condition without any visible damage. Replace every damaged and worn part.
2. The hose is connected correctly to the PAPR unit as well as to the headtop.
3. Airflow is sufficient (Section 4.2.).
4.2.
Airflow test
1. Disconnect the air hose from the unit.
2. Connect the flow indicator to the unit.
3. Switch the PAPR unit on and set the bottom airflow level.
If the pointer's top edge enters the red zone, airflow is insufficient and the filters must be replaced!
4.3.

Alarm test

To check if the alarm works properly, turn on the unit and close the air output with your palm. You should hear
and see alarm signals in less than 15 seconds.
5.

Maintenance, cleaning and decontamination

At the end of each working shift, it is recommended to check and clean all parts of the system and replace those
that are damaged.
Do not use aggressive detergents and solvents for cleaning! We recommend using common non-abrasive
cleaning products.
Cleaning products must never penetrate into the PAPR unit or the battery!
Use a damp cloth for cleaning and wipe the surface dry afterwards.
Cleaning should be done in a well ventilated room.
Beware of breathing in any particles and fibres settled on parts of the PAPR unit or its accessories!