CSAE 21UKEX2356X when installed as detailed in this
guide.
Declaration Requirements – EU Directive 2014/34/EU
This equipment is designed and manufactured to meet the
essential health and safety requirements not covered by
EU Type Examination Certificate SIRA 19ATEX2014X
when installed as detailed in this guide.
Pressure Safety
This equipment meets the requirements of the European
Pressure Equipment Directive 2014/68/EU and
UL 61010-1 for pressure safety.
Return Goods/Material Procedure
If the unit requires calibration or is unserviceable, return it
to the nearest Druck Service Centre listed at:
Druck.com/essential.
Do not dispose of this product as household waste. Use
an approved organization that collects and/or recycles
waste electrical and electronic equipment. For more
information, contact one of these:
- Our customer service department:
Druck.com/essential
- Your local government office.
Contact the Service Department to obtain a Return
Goods/Material Authorization (RGA or RMA). Provide the
following information for a RGA or RMA:
Product (i.e. DPI 705E / DPI 705E-IS)
Serial number.
Details of defect/work to be undertaken.
Calibration traceability requirements.
Operating conditions.

Quick Start

Key
Description
Switches instrument on or off.
Filter on or off.
Tare on or off (changes the display reading to zero).
Changes units (1 of 16 pressure units, °C or °F).
View maximum, minimum or perform leak test (Max/Min
is reset at switch on).
Back light. Short press illuminates for 5 seconds. Long
press toggles between permanent on and off.
Time Out
If a key is not pressed within 10 minutes, the instrument
times out and switches off.
Leak Test
To perform a leak test, press the
timer symbol
shows on the display with a leak test
duration of 60 seconds. Press the
duration to 120 or 300 seconds. To start the leak test, press
the
key again. The instrument counts down to zero,
then displays the leak rate. Press the
during the leak test to quit and return to normal
measurement.
Zero
A zero should be performed on gauge and differential
instruments before measuring pressure. To perform a zero,
open all pressure ports to atmospheric pressure. Press the
key three times. The
key to increase the
key at any time
and
keys together. The display briefly shows
and the instrument calculates a new zero.
Copyright 2019 Baker Hughes Company.
English–DPI 705E / 705E-IS Safety Manual | 3