Herunterladen Diese Seite drucken

aponorm FLEXIBLE ENTE Bedienungsanleitung Seite 2

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

THERMOMETER
FLEXIBLE ENTE
4
Cleaning and disinfecting area
art.-no. MT700
ENGLISH
1. Device overview
On/Off button
1
Display
2
Battery compartment cover
3
Measuring sensor
4
Flexible probe
5
2. Safety Instructions
Follow these instructions for use. This document contains important information
about the operation and safety of this device. Please read this document carefully before using
the device and keep it for future reference.
This device may only be used to measure human body temperature (in the mouth, rectum or
armpit). Do not attempt to use the device in any other measurement location, such as in the ear,
as this could lead to incorrect readings and/or injury.
Do not use the device if you notice any damage or if you notice anything unusual.
For personal hygiene, we recommend cleaning the appliance – before rst use –
according to the «Cleaning and disinfection» instructions below.
The minimum measurement period must be observed without exception until you hear the sig-
nal tone! Depending on the measurement location, it may be necessary to continue measuring
after the signal tone, see section «Measurement method/normal body temperature».
Make sure that children do not use the appliance unattended; some parts are so small that they
could be swallowed.
Do not use the device near strong electromagnetic elds such as mobile phones or radio equip-
ment. Keep a minimum distance of 3.3 m from such devices when using this device.
Protect the device from knocks and blows!
Avoid bending the thermometer measuring probe by more than 45°!
Avoid ambient temperatures above 60 °C. NEVER boil the appliance!
To clean the device, use only the disinfectants speci ed under «cleaning and disinfection».
We recommend that an accuracy test is carried out by an authorised laboratory for the instru-
ment every 2 years or after a mechanical impact (e.g. shock from falling).
ATTENTION: The measurement result displayed by this device is not a diagnosis! Do not
rely solely on the measurement result.
Batteries and electronic devices must not be disposed of in household waste but must be
disposed of in accordance with local regulations.
Protect from moisture.
Type BF application part
Study the manual carefully before use.
3. Preparing the thermometer for use
To switch on the thermometer, press the On/Off button
thermometer is switched on. First a display test is carried out – all segments should be display-
ed. Then, if the ambient temperature is less than 32 °C, an «---» ("low") and a ashing «°C»
appears on the display
. The thermometer is now ready for measuring the temperature.
2
4. Functional test
Every time the thermometer is switched on, a function test is carried out automatically. In the
event of a malfunction (inaccurate measurement), «ERR» appears in the display
device can no longer be used. Please contact your pharmacy for replacement (warranty check).
5. Using the thermometer
Select your preferred measuring method. During the measurement, the current temperature is
continuously displayed and the «°C» symbol ashes. If the signal tone sounds and the «°C» stops
ashing, this means that the measurement is nished, and the thermometer is ready for reading.
Depending on the measuring method, it could be necessary to continue the measurement longer than
is actual shows by the signal tone. Please refer «Measuring method/normal body temperature». The
thermometer can be switched off directly by brie y pressing the On/Off button
tery life. Otherwise the thermometer switches off automatically after approx. 10 minutes.
6. Display the last measured value
When turning on
the thermometer, the last measurement will be displayed with a «M» („Memory").
1
7. Measuring method/normal body temperature
7.1. Under the arm (axillary) / 34.7 - 37.3 °C (Adult) & 35.2 - 37.8 °C (Child)
Wipe the armpit with a dry cloth. Place the measuring sensor
the armpit as possible and clamp it to your body with your arm. This ensures that the room air
does not affect the measured value. Since the armpit takes longer to reach its stable temperatu-
re, wait at least 5 minutes regardless of the beep.
7.2. In the mouth cavity (oral) / 35.5 - 37.5 °C (Adult) & 36.0 - 38.2 °C (Child)
Do not eat or drink anything warm or very cold 10 minutes before the measurement. The mouth
should be closed at least 2 minutes before starting a measurement. Place the thermometer with
the measuring sensor
into one of the two tongue pockets, directly at the root of the tongue.
4
The measuring sensor should be in good contact with the tissue. Press your lips rmly together
and hold your mouth closed, so that the measurement result is not in uenced by the air you are
breathing. Breathe calmly through your nose during the entire measurement process. If this is
not possible due to obstructed airways, another measuring method should be used.
Minimum measurement time: 30 seconds!
Vertrieb:
Hersteller
WEPA Apothekenbedarf
Microlife AG
GmbH & Co KG
Espenstrasse 139
D-56204 Hillscheid
9443 Widnau / Switzerland
www.wepa-dieapothekenmarke.de
www.microlife.com
PZN-10545841
Bestell-Nr. 047234
Measuring sensor
Flexible probe
5
; a long signal tone indicates that the
1
2
– this extends bat-
1
*
of the thermometer as far up in
4
Microlife UAB
P. Lukšio g. 32
08222 Vilnius
Lithuania
2
Display
1
7.3. In the anus (rectal) / 36.6 - 38.0 °C (Adult) & 37.0 - 38.4 °C (Child)
For rectal measurement, it is best to lie on your back or on your side and bend your legs. To
avoid injuries, it is recommended to additionally grease the measuring tip (e.g. with Vaseline).
The measuring sensor
If you are not sure about this method of measurement, you should consult a specialist for inst-
ruction/training.
Minimum measurement time: 30 seconds!
Don't be too strict about keeping within the exact normal temperature ranges – rather consider
*
them as guidelines that should be in line with your physical condition.
8. Cleaning and disinfection
For disinfecting the thermometer at home, use an isopropyl alcohol swab (70%) or a cotton cloth
moistened with 70% isopropyl alcohol to wipe off coarse, visible surface dirt from the thermo-
meter measuring probe
on and safety instructions of the disinfectant manufacturer. Then immerse the thermometer
measuring probe
nutes (max. 24 hours). Let the disinfectant dry for 1 minute before using the thermometer again.
Avoid immersing or wiping the display
9. Battery replacement
As soon as the «» symbol appears on the right of the display
it
must
be
compartment
and replace the exhausted battery with the «+» sign pointing upwards. Make
3
sure that the same type of battery is used. Batteries are available in every electrical shop. For
disposal, please observe the instructions under «Safety instructions».
10. Technical speci cation
Type
Measuring range
Measurement accuracy
Operating conditions
Self-test
Display
Memory
Storage conditions
Battery
Battery life
IP class
Standard
Electromagnetic
compatibility
Average lifetime
. In this case the
11. Guarantee
This device is guaranteed for 30 years from the date of purchase. The guarantee is only valid if
a sales receipt issued by the dealer is presented. Mailing costs for sending the thermometer in
by the customer will not be reimbursed. For customers outside Germany, the processing of the
guarantee case also depends on the reimbursement in advance of the costs required for the re-
turn shipment.
The legal rights of the consumer (e.g. warranty rights) are not limited by this guarantee and re-
main unaffected by it.
The guarantee only covers the thermometer itself. Batteries and packaging are not covered by
the guarantee.
If the device is opened or if changes are made to it, the guarantee becomes void.
The guarantee does not cover damage caused by improper handling, discharged batteries,
accidents or failure to follow the instructions for use.
Please always contact your specialist dealer/pharmacy or our aponorm
(info@aponorm.de)
You can nd detailed information about our thermometers, blood pressure monitors, inhalers and
our services at www.aponorm.de
Date of purchase:
Serious incidents must be reported to the manufacturer and the proper authority of the member
state in which the user and/or patient is established.
Stand der
Information: 02/2021
3
Battery
compartment cover
On/Off button
of the thermometer is carefully inserted 2-3 cm into the rectum/anus.
4
(= complete cross-hatched area). Note: Please observe the applicati-
5
(= complete cross-hatched area) in 70% isopropyl alcohol for at least 5 mi-
5
to protect it from damage.
2
changed.
To
do
this,
remove
Maximum thermometer
32.0 °C to 43.9 °C
Temp. < 32.0 °C: Display «---» for "low" (too low)
Temp. > 43.9 °C: Display «H» for "high" (too high)
± 0.1 °C between 34.0 to 42.0 °C
± 0.2 °C between 32.0 to 33.9 °C
and 42.1 to 43.9 °C
10.0 °C to 40.0 °C
Automatic internal check for 37.0 °C test value;
«ERR» (error) is displayed if the deviation is > 0.1 °C
Liquid crystal display (LCD) with three digits; smallest display
unit = 0.1 °C
For automatically storing the value measured last
Temperature: - 25.0 °C to + 60.0 °C,
Air humidity: 15% to 95% (without condensation)
LR41 (1.5V) / SR41 (1.55V)
approx. 4,500 measurements (with new battery)
IP22
EN 12470-3, clinical thermometer; ASTM E1112;
IEC 60601-1; IEC 60601-1-2 (EMC); IEC 60601-1-11
Device complies with the speci cations of the IEC 60601-1-2
standard
5 years or 10,000 measurements
, the battery is exhausted, and
2
the
cover
of
the
battery
customer service
®
Medizinprodukt