Herunterladen Inhalt Inhalt Diese Seite drucken

Use Of The Table; Maintenance - LEMI GEMYA Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

200 kg (max. patient weight)
On the table, only one person can get up; the person must go up/down when the table is in the basic position
going up on the seat.
WARNING! Stretch out on the table always sitting properly in
the central part of the seat.
Do not sit or lean on the side of the feet or of the head.
WARNING! The limbs of the patient must always be on the table. The operator must
always inform the patient when he moves the table.
WARNING! Be careful to make sure that the table or its electrical parts are not wet
and that liquids become not stagnant on the table itself; immediately wipe up any li-
quids – including physiological liquids of the patient – which may end up on the table.
To avoid shock hazard, do not connect or disconnect the power plug with wet hands.
Do not use the table barefoot.
The table has been designed and constructed so as to reduce as much as possible the necessary maintenance
and ensure proper functioning over the years without any problems.
During any intervention on the table, it is recommended to follow the instructions given in this section.
WARNING! All interventions on the table must be carried out by competent and trai-
ned personnel on the methods to operate correctly and safety.
Do not make any action on the table moving, and always check that the plug con-
necting to power supply is disconnected; bring the power plug next to you during the
execution of the intervention on the table so you can check at any time the actual
separation.

8. USE OF THE TABLE

2500 N (Safety load)

9. MAINTENANCE

13

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Gemya evoPortofinoPortofino evo

Inhaltsverzeichnis