Effective weight calibration
+
3 sec.
CAlibr
PArAM
CAL
Set sample weight
During the calibration phase, the display shows the weight intermittently.
Zero calibration
Full scale calibration
Set password
In out
SEriAl
AnAloG
Zero calibration
FS calibration
Perform the operation on unloaded scale but complete with tare and sta-
bilized weight. The displayed weight must be reset. This operation can be
repeated several times
Before carrying out the operation, load the sample weight on the scale
and wait for stabilization; the display shows the detected value to be cali-
brated. If the value set is higher than the resolution offered by the instru-
ment, it is not accepted and the display shows an error message for a
few seconds. It is always possible to repeat the calibration operations.
Pagina –
page
–
opTion
Const
Seite
68
Exit menu