Typical Measurements and Accuracy (cont'd)
Resistance
Measurement
Resistance
Range
600.0 Ω
± (1.0% + 0.5 Ω)
6.000 kΩ
± (1.0% + 0.005 kΩ)
60.00 kΩ
± (1.0% + 0.05 kΩ)
Open Circuit Voltage (typical): 1.7 VDC
Capacitance
Measurement
Capacitance
Range
200.0 μF
± (2.0% + 0.4 μF)
2500 μF
± (2.0% + 4 μF)
* For film capacitors (capacitors with
negligible dielectric absorption).
14
SELECT
Accuracy
Accuracy*
Continuity Check
Continuity
Tone Threshold: Between 10 Ω and 250 Ω
Response Time: Approximately 32 ms
Diode
Measurement
Reverse Bias
Forward Bias
Diode Test
Measuring Range: 2.000 V
Accuracy: ± (1.5% + 0.005 V)
Test Current (typical): 0.3 mA
Open Circuit Voltage (typical):
< 3.5 VDC