(c) Cycle the power off and then back on.
the console will now be in "Pressure
feedback" mode. The display will now
indicate a representation of in-joint
pressure once a tubing set has been
loaded.
(d) To display the pressure set-
point, press the Pressure Up
or Down button once. The
display will indicate the
current set pressure for
approximately three (3)
seconds and then revert back to a
representation of in-joint pressure.
This procedure need only be performed once, as
the console retains this mode in memory, even if
power is turned off or interrupted.
2.
To change the display to the "Set Point"
mode, first ensure that there is not a tubing
set loaded. Ensure the power is off and
perform the following steps:
(a) Press and hold the Pressure
Down, Pressure Up, and
"Run/Stop" buttons
simultaneously while turning
the console power on. Release
all buttons once the display
indicates"d 1".
(b) Place a finger on the face of the pump
to the left of the remote control
receptacle, and to the right of the
pressure sensor window. While
applying slight pressure, slide your
finger toward the pressure sensor
window until the display indicates
"Poi".
(c) Cycle the power off and then back on.
The console will now be in "Set Point"
mode.
E-13