(e) Circulator - Close the bag clamps.
(f) Circulator - Hang the fluid bag(s) on
the infusion pole, so that the bottom of
the bag(s) is at a minimum of 12 inches
from the top of the pump, and spike the
bag(s).
6.
Prime the Tubing Set:
Using the 10k100 and 10k125
(a) Scrub - Ensure the inflow clamp is
open. Hold the patient end of the
tubing over a basin.
(b) Circulator - Open the bag clamps.
(c) Circulator - Press the "Run/
Stop" button on the pump or
the remote control.
(d) Scrub - After all air is purged
from the tubing, close the
patient inflow clamp.
(e) Scrub - Attach the luer fitting on the
Tubing Set to the inflow device.
Using the 10k150
(a) Scrub - Ensure the inflow clamp is
open. Hold the patient end of the
tubing over a basin.
(b) Circulator - Open the bag clamps.
(c) Circulator - Press the "Run/
Stop" button on the pump or
the remote control.
(d) Scrub - After all air is purged
from the tubing, close the
patient inflow clamp.
(e) Scrub - Attach the luer fitting on the
Tubing Set to the inflow device.
(f) Scrub - Separate the outflow line from
the inflow line, as needed, by pulling
the two lines apart. Connect the
outflow tubing line to the outflow
device or shaver.
(g) Circulator - Connect the suction/
gravity tubing line to the appropriate
hospital discharge mechanism.
Using the 10k200
(a) Scrub - Direct the distal tip of the
cannula towards the basin and depress
and hold the blue inflow button.
(b) Circulator - Open the bag clamps.
(c) Circulator - Press the "Run/
Stop" button on the pump or
the remote control.
(d) Scrub - After all air is purged
from the tubing, release the
blue inflow button.
(e) Circulator - Connect the suction/
gravity tubing line to the appropriate
hospital discharge mechanism.
7. The pump is now ready for use.
8. To adjust irrigation pressure:
(a) Press the appropriate button
to increase or decrease the