ilization, inspect the device in the autoclave bag
for integrity. If suspicion about the item exists,
discard it and order a replacement from Parkell
or your dealer.
• DO NOT USE DRY HEAT OR CHEMCLAVE on
the sheath or the insert.
• Detachable handpiece sheath and ultrasonic
inserts from Parkell are designed to withstand a
minimum of 250 autoclaving cycles when repro-
cessed as per the recommended autoclaving
instructions by the manufacturer.
How to Replace the Filter Disk
The water filter disk coming off of the rear hose of
the unit should be replaced when it becomes dark
or clogged with debris so that water flow is not
blocked (1-3 months). Replacement Filters are
available from Parkell (hD419 – 10 per package).
1. Close any water valves, and bleed all water lines
of residual water pressure.
2. Disconnect the scaler water supply hose from
the water supply at the quick- connect.
3. Remove the old filter disk by holding it in one
hand and loosening it by twisting it off the hose
connectors on either side. They rotate in oppo-
site directions.
4. Install the new filter disk in the same position
as the old one, and tighten the connectors in
reverse.
5. Reconnect the scaler water supply hose to the
water supply via the quick-connect and check
for leaks.
6. Note: If the filter becomes clogged and you do
not have a replacement disk, the hose connec-
tors may be locked together TEMPORARILY
without the filter disk to allow emergency opera-
tion only.

Troubleshooting

Unit does not operate
("power on" indicator does not light):
• Transformer cable disconnected from wall
outlet or scaler. Reconnect properly.
• Wall outlet not live. Reconnect properly.
• Foot Pedal or cable is damaged.
Contact Parkell.
• Unit fuse has failed. Contact Parkell.
Insert does not vibrate properly:
• Faulty, damaged or worn insert. Replace.
• Insert not correctly seated in handpiece.
Reinsert properly.
• Use a new Parkell
• Power control not correctly adjusted.
Readjust.
Excess heat at handle or tip of insert:
• Faulty, damaged or worn insert. Replace.
• Too little water flow. Insert may be blocked.
Inspect and clear blockage, or replace insert.
• Power too high for the amount of water.
Increase water flow.
• Excessive hand pressure applied. Correct
technique.
No water spray when foot pedal is
depressed:
• Water line blocked or kinked. Correct.
• Water passage in the scaling insert clogged.
Inspect and clear blockage, or replace insert.
• Water filter clogged. Replace.
• No water supply connected. Correct.
Insert does not go into handpiece:
• Insert stack is bent. Straighten plates
carefully by hand and reinsert.
• O-ring is too tight. Lubricate O-ring with water
and use twisting motion to seat insert.
• Autoclavable handpiece sheath is not
attached to handpiece.
Insert falls out of handpiece or water leaks
from front of handpiece:
• O-ring damaged or worn. Replace O-ring on
insert and/or handpiece.
Water leaks at base of handpiece:
• Orange O-ring at end of handpiece is missing
or damaged.
Insert comes out of handpiece when pedal
is depressed:
• Water pressure is too high. Adjust water
pressure to 15-35 psi at dental unit connec-
tion or in floor junction box, or replace insert
O-ring.
User Maintenance and
Authorized Service
• The Parkell TurboSENSOR requires very little
maintenance for years of trouble-free operation.
There are no user-serviceable parts inside the
control unit or the foot pedal. Opening the cases
will void all warranties. Authorized Parkell person-
nel must make all internal repairs.