Rx Only (product intended for use by a Licensed dental professional). This manual, in whole or in part,
should not be considered a substitute for formal training in ultrasonic scaling. Appropriate professional
education is STRONGLY RECOMMENDED prior to using this device clinically.
Instructions for Use
Device Description
The Parkell TurboSENSOR
tostrictive ultrasonic tooth scaler that operates at
either 25 KHz or 30 KHz. The TurboSENSOR auto-
matically detects whether the handpiece contains
a 25 KHz or 30 KHz insert, and switches to the
correct operating frequency. The TurboSENSOR
has an expanded low-power range to improve
patient comfort during debridement, and a
power-boosting, user-controlled Turbo feature to
increase scaling power when needed for heavy
deposits. The unit is designed to work with Parkell
inserts. For information about usage with non-Par-
kell inserts, please see our warranty policy.
Indications/Intended for Uses
For removal of calculus, plaque and oral debris
from teeth during dental prophylaxis/scaling.
Contraindications
Because of the potential for electromagnetic inter-
ference, this device should not be used on patients
or by clinicians with cardiac pacemakers, internal
defi brillators, intracorporeal fl uid pumps or any
other implantable electronic devices, or in close
proximity to sensitive patient monitoring devices
such as pulse oximeters.
25KHz or
30KHz Inserts:
Device automatically
switches operating
frequency to match
(insert not included).
Water Control Knob
is an autotune magne-
®
"PERIO" Light:
Indicates low-power range where
large changes in knob produce
subtle changes in tip amplitude.
Ultrasonic Scaler ( hD560-110, hD560-230)
If patient or operator is pregnant, or has any medi-
cal condition which might be affected by this device
during treatment, consult a physician prior to use.
In the event of any adverse reaction from the
patient, discontinue the scaling procedure.
Warnings
• The water supply to the scaler should always be
turned off whenever the device is being
connected, disconnected, or when not in use.
• The power supply cord functions as the AC power
disconnect device for the scaler.
• The transformer should be located above and
away from any sources of water that may enter the
unit. As with all electrical devices, the unit should
not be immersed in water or other liquids. Do not
reach for the device if it has fallen into liquid until
power is disconnected, and do not use the device
after it has fallen into liquid. Return the device to
Parkell for servicing.
• Portable RF communications equipment (includ-
ing peripherals such as antenna cables and exter-
nal antennas) should be used no closer than 30
cm (12 inches) to any part of the [ME EQUIPMENT
or ME SYSTEM], including cables specifi ed by the
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
1
A00414revO0121
Autoclavable
Removable
Handpiece Sheath.
"ON" Indicator:
Lights only when
footswitch is
depressed.
"TURBO" Light:
Indicates the device is operating
in high-power "Turbo" mode
(footswitch fully depressed).
EN
®

Werbung

Inhaltsverzeichnis
loading

Inhaltszusammenfassung für parkell TurboSENSOR

  • Seite 1 ® during treatment, consult a physician prior to use. tostrictive ultrasonic tooth scaler that operates at either 25 KHz or 30 KHz. The TurboSENSOR auto- In the event of any adverse reaction from the matically detects whether the handpiece contains patient, discontinue the scaling procedure.
  • Seite 2: Specifications

    • Keep the long axis of the insert tip parallel to included with the basic scaler unit. The unit is the long axis of the tooth to wipe deposits from designed to work with Parkell inserts. For informa- the tooth. tion about usage with non-Parkell inserts, please see our warranty policy.
  • Seite 3: Electrical Connection

    Adec-type or other when the device is not in use. type of connection, remove The TurboSENSOR gives you two ways to adjust the old fitting from the unit the scaling power, by using the Power Control Knob hose and attach it to the for normal operation, or by depressing the Foot new water line.
  • Seite 4 NEVER OPERATE UNIT Dental Handpieces”. WITHOUT WATER FLOWING. • DO NOT AUTOCLAVE the TurboSENSOR • Use the lowest effective scaling power for control unit, as autoclave heat will destroy it. The the case at hand. This keeps heat generation to control unit should be cleaned of debris with a a minimum.
  • Seite 5: Troubleshooting

    Insert does not vibrate properly: for integrity. If suspicion about the item exists, • Faulty, damaged or worn insert. Replace. discard it and order a replacement from Parkell • Insert not correctly seated in handpiece. or your dealer. Reinsert properly.
  • Seite 6: Explanation Of Symbols Used

    TurboSENSOR, call replacement parts. Technical Support Service at 1-800-243-7446, M-F • Within the US, all repairs must be made by Parkell. from 8:30 AM to 5:00 PM EST. We maintain a complete service and parts facility...
  • Seite 7: Gerätebeschreibung

    Anwender per Fußschalter gesteuert werden, um die strom-Trenneinrichtung für den Scaler. Leistung beim Scaling bei Bedarf zu erhöhen. Das Gerät ist für den Betrieb mit Einsätzen von Parkell vorgesehen. • Der Transformator sollte oberhalb und in ausreichen- Informationen über die Verwendung in Kombination mit dem Abstand von allen Wasserquellen, aus denen unter Einsätzen anderer Anbieter oder Hersteller entnehmen...
  • Seite 8: Spezifikationen

    Scaler ist der Zugang zu einer geerdeten Steckdose und NORMENKONFORMITÄT einem Wasseranschluss mit Trinkwasserqualität erforder- Der Parkell TurboSENSOR ist TÜV-geprüft und konform lich. Das Gerät und sein separater Transformator erzeugen mit den Normen IEC 60601-1, 60601-1-2 und CAN/ eine minimale Menge an Abwärme. Sie dürfen nicht abge- CSA C22.2 Nr.
  • Seite 9: Elektrischer Anschluss

    Fußpedal auf die erste Position gedrückt wird, fängt der Der Scaler darf nur über das Transformatoraggregat, mit Einsatz an, mit mittlerer Amplitude zu vibrieren. dem das Gerät von Parkell geliefert wird, mit Strom versorgt werden. • Wenn der Leistungsreglerknopf auf die Einstellung „Perio“...
  • Seite 10: Reinigung Und Infektionsprävention

    Autoklavierbeutel auf Integrität. Falls diesbezüglich • Übermäßige Druckanwendung mit der Hand. Zweifel bei dem Artikel besteht, verwerfen Sie ihn und bestellen Sie einen Ersatz von Parkell oder Ihrem • Nichteinhalten der niedrigsten wirksamen Leistungsein- Händler. stellung bei der bevorstehenden Behandlung.
  • Seite 11: Problembehandlung

    Sie Parkell. • Gerätesicherung ist ausgefallen. Kontaktieren Sie WARTUNG DURCH DEN ANWENDER UND Parkell. AUTORISIERTE REPARATUREN • Für den Parkell TurboSENSOR sind über Jahre des ■ Einsatz vibriert nicht ordnungsgemäß: störungsfreien Betriebs nur sehr wenige Wartungsar- • Fehlerhafter, beschädigter oder abgenutzter beiten erforderlich.
  • Seite 12: Garantie Und Nutzungsbedingungen

    Unternehmen repariert: Parkell, Inc., 300 Executive Drive, Edgewood, NY 11717 Telefon (gebührenfrei): (800) 243-7446 • Telefon: (631) 249-1134 • www.parkell.com This precision dental device was designed, manufactured and is serviced in the United States of America by: 300 Executive Drive, Edgewood, NY 11717 Toll-Free: (800) 243-7446 •...

Inhaltsverzeichnis