9. Cleaning and sterilization

Method of sterilization:
Put this product in a sterilization pack (or foil) and place it on a sterilization tray for autoclave sterilization with reference to
the following terms:
1.Sterilize with steam at 134°C (273°F) / 2,1 bars / 18 min.for endodontic instruments and filling.
2.Do not use high pressure steam sterilizer which heats more than 200 degrees Celsius including drying process.
3.When reusing the instrument, wash away foreign substances completely and sterilize.
4.Regarding use of medical cleaning agent, follow the instruction manual by its manufacturer strictly.
5. Dispose the product if damaged or decreased in performance.
10. Warnings
a. Only skilled dentists are allowed to use.
b. Be sure to sterilize this product for each use.
c. Do not use this product except for the dental service and treatment. Use it in accordance with the intended use.
d. Please confirm that patient has no allergy to Nickle Titanium alloy and other medical reaction before the treatment.
e. These instruments are intended for use only in dental root canal treatment, users must follow a correct and reasonable
way under medical environments.
f. Please use a rubber dam system to ensure patients' safety when deliver endodontic treatment.
g. Do not use this product for a patient who indicates sensitization and allergic reaction.
11. Storages and duration of use
a. Avoid storing at high temperature, humidity and direct sunlight. Keep liquids away. Store it at room temperature.
b. Do not damage or make a pinhole to packaging materials.
c. This product is subject to be improved without previous notice. Enforce first-in first-out method for stock management.
d. Do not store the instrument under a germicidal lamp to avoid deterioration.
Mouvement alternatif
Alternating movement
6 pièces par boîte
6 pcs/box
Produit vendu non-stérile
Non sterile product
Pour l'utilisation des moteurs endo
For endo motor use
Wellkang Ltd. Suite B,
29 Harley Street London, W1G 9QR,
England, UNITED KINGDOM
DESIGNED IN FRANCE
Shenzhen SuperLine
Technology Co. Ltd.
0 1 9 7
Room 1206, Building 4 Chongwen Garden,
Taoyuan Street Nanshan District
Shenzhen 518055 Guangdong China
CADENCE
IMP
2 bis, chemin du Loup
93290 Tremblay-en-France - FRANCE
N'est pas retournable si la boîte a été
ouverte-Non returnable if seal is broken
Produit en nickel-titane
Nickel titanium product
Les stoppers sont en silicone
Stoppers are silicon stoppers
Stériliser à 134°C
Sterilize at 134° C
Version 1 – 11/2020

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Access EASY-FILE FLEX Serie

Inhaltsverzeichnis