EMC data
Guidance and manufacturer´s declaration – electromagnetic emission - for all EQUIPMENT
AND SYSTEMS
The model LX260TE is intended for use in the electromagnetic environments specified below. the
customer or the user of the model LX-260TE should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /flicker
emissions IEC 61000-3-3
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrostatic
transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
54
Compliance
Group 1
Class B
N/A
N/A
IEC 60601
test level
± 8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15
kV air
± 2 kV for power supply
lines 100 kHz repetition
frequency
±1 kV for input/output
lines
± 0.5 kV, ± 1 kV dif-
ferential
mode
line-line
IFU_G01079_ML_ V02202006R02
The Model LX-260TE uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfer-
ence in nearby electronic equipment.
The Model LX-260TE is suitable for use in all
establishments, including domestic establish-
ments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Compliance level
Floors should be wood, concrete
± 8 kV contact
or ceramic tile. If floors are covered
±2 kV, ±4 kV, ±8
with synthetic material, the relative
kV, ±15 kV air
humidity should be at least 30 %.
N/A
N/A
Electromagnetic
environment – guidance
Electromagnetic
environment – guidance
N/A
N/A