Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

This method allows you to locate the centre of the acupuncture point relatively accurately.
To search for another point, first remove the 3B LASER from the skin. The laser switches to the automatic status
again, the measured value remains stored.
To delete the stored value you have to press the SET key. After doing that you are back in the basic setting and the
point search modus is OFF.
The complete procedure as described must be repeated each time you search for a new point.
Please note:
Switching-on the laser does not necessarily mean that you have found an acupuncture point. It only means that the
current conductivity value is higher than the stored value. It may well be the case that you have to carry out several
measurements before you find the point.
Working with the 3B LASER does not replace knowledge of the positions of acupuncture points. Extremely moist,
dry or hairy skin, or also an extremely high atmospheric humidity in the room, may lead to faulty results.
Acupuncture points can possibly lie very deep under the skin and are extremely difficult to find. The same applies
to points which are not active. It is extremely difficult, and sometimes even impossible, to find such points using
a point searcher. A measurement is very accurate if the acupuncture points are located relatively far apart. It is un-
derstandable that point searchers will have difficulties if the acupuncture points are situated very closely together.

Application

The following must be observed if you wish to achieve optimum
success with the therapy.
• To keep reflection losses as low as possible, skin must be
grease-free, and the 3B LASER must be held perpendicular
to the skin.
• A general rule that applies to all treatments is the principle
of tapering dosage. Initial treatments should last no longer
than 2 minutes.
• Maximum treatment time per patient is 20 min. After this time the unit turns off automatically.
• The distance between the applicator and the skin should be no more than a few mm, or direct contact.
28
Downloaded from
www.Manualslib.com
manuals search engine
measurement (A3)
value e. g. 90
measurement (A2)
value e. g. 80
measurement (A1)
value e. g. 60
Last revised 16.02.12

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für 3B LASER PEN 200 mW

Inhaltsverzeichnis