confirm the alarm on the device. Without alarm confirmation, the device assumes that the root
cause has not been solved and will not raise a new alarm. The device does not document who
confirmed the alarm. If the user needs this information, the documentation is responsibility of
customer, including limiting the physical access to the device.
• The Fridge-tag 2 L carries a unique identifier. The user needs to document which refrigerator/
freezer the Fridge-tag 2 L is monitoring in which time span, if needed for the user to comply with
regulation. Additionally, the user needs to document the unique identifier of the temperature
sensor used and its calibration certificate.
• In order to have a sound administration (as required by regulation), the user needs to ensure that
the date and time on the Fridge-tag 2 L is correct. It is strongly recommended that the UTC time
zone is used, to prevent issues with daylight saving time. However, when another time zone is
used, the manufacturer recommends documentation (in SOP) when the date/time on the Fridge-
tag 2 L is adjusted and by whom. It is the user's responsibility to ensure that data of Fridge-tag 2 L
can be correlated with other information, such as inventory kept in refrigerator/freezer.
• Before digital reports of Fridge-tag 2 L are used, the user must validate their authenticity using the
Verifier. The Verifier can be obtained on Berlinger website.
Other considerations
• The Fridge-tag 2 L has no notion of what correct storage conditions are of the products the user
stores in refrigerator/freezer. The user is assumed to be aware of what is correct and must be able
to assess the impact of measured temperature in a refrigerator/freezer on his products.
• The Fridge-tag 2 L has no notion of what the user stores in the refrigerator/freezer, hence the user
must document and manage that himself. Additionally, the correlation between measured
temperature and the product(s) must be done by the user himself.
• Information generated by Fridge-tag 2 L is lost, unavailable or unreliable:
- When the Fridge-tag 2 L is damaged, defective, is used beyond its expiration date or indicates
'empty battery'.
- When external sensor is not attached or is defective.
- When the (digital) report is not archived before older data is overwritten. The maximum report
length (counted from the current moment) is defined. The Fridge-tag 2 L does not warn when older
information is overwritten.
- When the calibration of the temperature sensor has expired.
• The Fridge-tag 2 L is compatible with Berlinger SmartView. When the user has a Berlinger
SmartView license, the information of Fridge-tag 2 L must be uploaded manually to Berlinger
SmartView. Berlinger SmartView assumes that the time and date have been set correctly on
Fridge-tag 2 L. When Berlinger SmartView is used to analyze recorded temperature data of
Fridge-tag 2 L, it may present different duration values than the Fridge-tag 2 L itself. This is due to
the fact that the devices measure and analyses temperature on a minute interval, while Berlinger
SmartView analyses the data on (given) logging interval. The difference in event duration can be
up to two (2) times the logging interval.
• Berlinger gives no guarantees that the Fridge-tag 2 L is compatible with third party servers or
systems, or that all functions of the device work when used with third party systems. Even if at
moment of purchase the device is compatible, no guarantee is given for future.