Herunterladen Inhalt Inhalt Diese Seite drucken

Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Becton Dickinson 5513E Bedienungsanleitung

Surgical clipper and charging adapter stand
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
REF 5513E/5514 series is intended for use in the electromagnetic environment specified below�
The customer or the user of REF 5513E/5514 series should insure that it is used in such an environment�
Immunity test
IEC 60601 test level
3V
0�15MHz-80MHz
Conducted RF
6V in ISM bands
IEC 61000-4-6
between 0�15MHz and
80MHz
80% AM at 1 kHz
3V/m
Radiated RF
80MHz-2,7GHz
IEC 61000-4-3
80% AM at 1 kHz
Proximity fields
from RF wireless
IEC 60601-2: 2014
communications
Table 9
equipment
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies�
NOTE 2: These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people�
12
Guidance and manufacturer's declaration — electromagnetic immunity
Compliance level
3V
0�15MHz-80MHz
6V in ISM and
amateur radio bands
between 0�15MHz and
80MHz
80% AM at 1 kHz
10V/m
80MHz-2,7GHz
80% AM at 1 kHz
IEC 60601-1-2: 2014
Table 9
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no
closer to any part of REF 5513E/5514 series, including cables, than the
recommended separation distance calculated from the equation applicable to
the frequency of the transmitter�
Recommended separation distance
d = 1.2 √P
d = 0.35√P 80MHz to 800MHz
d = 0.7√P 800MHz to 2�7GHz
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in meters (m)�
Field strength from fixed RF transmitters, as determined by an
electromagnetic site survey,
should be less than the compliance level in
a
each frequency range�
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
d = 6/E√P
380MHz to 5�8GHz

Werbung