length (with lightguide) - 196 mm
4. Weight:
– Charging station - 560 g
– Cordless handpiece (with battery & lightguide) - 245 g
5. Light intensity – max. 1300 mW/sq.cm.
6. emitting time - 10 - 20 - 30 sec. / 9 min. /±5%/;
7. emitted light - blue, 430 - 490 nm
8. Light source - LeD module
9. curing modes - LonG, raMP, fast, PuLse.
10. audible signal every 10 sec. / 1 min.
11. Possibility for start and stop of emitting handpiece at any moment.
12. Digital display of curing mode and countdown timer.
13. Battery - Li-Ion 2400 mah, 3,7 V
14. Low battery indication.
15. Working mode (1 min. / 10 min.)
the manufacturer of this unit declares to provide on request all additional necessary technical documentation / information which will help the
user's technical staff to service the parts of the unit which the manufacturer has claimed to be a subject for repair.
III. ACCESSORIES AND SPARES
1. charging station - 1 pce.
2. cordless handpiece - 1 pce.
3. Battery - 1 pce.
4. adapter - 1 pce.
5. Lightguide - 1 pce.
6. Protection shield - 1 pce.
7. operations guide - 1 pce.
IV. SAFETY PRECAUTIONS
as with all technical devices, the proper function and safe operation of this unit depend on the user's compliance with the standard safety pro-
cedures as well as the specific safety recommendations described in this operations guide.
1. the unit must be used in strict accordance with the operations guide. the manufacturer accepts no liability for any
damage resulting from the use of this unit for any other purposes.
2. Before start-up of the unit make sure that the operating voltage indicated on the unit label corresponds to the mains voltage. operation of
the unit at a different voltage may damage the unit. electrical safety is ensured by Protection class II against electrical current damages accor-
ding to en 60601-1. use of "shuko" plug or unit's grounding is not necessary.
M+W Superlite 1300 should be operated in indoor premises, in the following conditions:
• temperature +10° - +40°;
• relative air humidity -30 - 75%;
• atm. pressure 700 - 1060 hPa;
• absence of chemically active or flammable substances.
3. Irradiation of the eyes bears certain risk for their damage. therefore the light must not be directed toward the eyes! exposure must be re-
stricted to the working area. use of protection glasses and/or protection shield that blocks effectively the blue spectral range is highly recom-
mended!
ATTENTION! the emitted light is in the wavelength range 430-490 nanometers. During work use protection goggles and/or protection
shield!
4. the unit can be used only doctor's advice by persons with photobiological reactions; persons currently taking photosensitive medication; per-
sons with cataract surgery; persons with retinal diseases, etc.
5. solvents, flammable liquids and powerful heat sources must not be kept near the unit as exposure to one of these
8