2. Disinfect the handpiece with a soft cloth and disinfectant solution. Please do not spray, only wipe. Liquids should not penetrate into the in-
terior of the handpiece. Do not use abrasive cleaning agents. We recommend M+W alcohol-free disinfection spray. Moisten the cloth and ca-
refully wipe the handpiece.
remove and remaining composite residue from the light conductor.
Manual and machine preparation of light conductor, glare protection shield and light protection cap:
- do not use abrasive cleaning agents.or wire brushes.
Manual cleaning
accessories: brushes with soft bristles, lint-free cloths, alkaline cleaning agents, i.e. M+W Instrument bath af, distilled water
step
Duration
1
1 min.
2
3
1 min.
4
1 min.
5
Machine preparation using a CDD (only light conductor and light protection cap)
Preparation: for pre-cleaning see step 1 for manual cleaning. Visual inspection. should visible contamination persist, pre-clean again manually
until the contamination is removed.
then place in the thermal disinfector. ensure that the light conductors cannot come into contact with other objects during preparation (risk of
breakage!)
start the cleaning programme of the thermal disinfector:
Programme
Duration
Pre-cleaning
2 Min.
cleaning
2 Min.
neutralisation
5 Min.
Disinfection
5 Min.
rinsing
2 Min.
Drying 40 Min.
90° c
After cleaning is completed, check the parts carefully for damage, scratches, etc. Dispose of damaged parts.
Sterilisation of lighting conductor and light protection cap:
Pack parts in suitable packaging material (according to DIn 58 953/7 or en 868)
then sterilise in autoclave at 134°c
VII. WARRANTY CONDITIONS
1. Warranty period of the M+W superlite 1300 curing light is 24 months from the date of purchase. If date of purchase is not written, warranty
starts at date of production.
2. During the warranty period the replacement of the defective parts (except for lightguide; LeD module and battery -
6 months warranty) is done free of charge by manufacturer's service.
3. If during the guarantee period the unit is damaged, due to incorrect operation (mechanical, chemical, thermal or electrical damages), da-
Description
rinse used parts under cold water and remove coarse, visible contamination
immediately.
now place pre-cleaned parts in alkaline cleaning agent. Please ensure complete
wetting. observe application concentration and exposure time of the disinfectant
agent.
then rinse with clean water.should visible contamination persist,
please repeat steps 1-3 until all visible contamination has been removed.
then rinse parts with distilled water and dry.
final visual inspection for contamination. If necessary, repeat steps 1-6.