Herunterladen Inhalt Inhalt Diese Seite drucken

Declaration Of Conformity; General Information Regarding The Ce Marking - Hameg HM8001-2 Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für HM8001-2:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

G e n e r a l i n f o r m a t i o n
DECLARATION OF
CONFORMITY
HAMEG Instruments GmbH
Industriestraße 6
D-63533 Mainhausen
The HAMEG Instruments GmbH herewith
declares conformity of the product
Product name: Main frame
Type:
HM8001-2
with:
Options:
HO801
complies with the provisions of the Directive
of the Council of the European Union on the
approximation of the laws of the Member States
z relating to electrical equipment for use within
defined voltage limits (2006/95/EC) [LVD]
z relating to electromagnetic compatibility
(2004/108/EC) [EMCD]
z relating to restriction of the use of hazardous
substances in electrical and electronic
equipment (2011/65/EC) [RoHS].
Conformity with LVD and EMCD is proven by
compliance with the following standards:
EN 61010-1: 04/2015
EN 61326-1: 07/2013
EN 55011: 11/2014
EN 61000-4-2: 12/2009
EN 61000-4-3: 04/2011
EN 61000-4-4: 04/2013
EN 61000-4-5: 03/2015
EN 61000-4-6: 08/2014
EN 61000-4-11: 02/2005
For the assessment of electromagnetic
compatibility, the limits of radio interference for
Class B equipment as well as the immunity to
interference for operation in industry have been
used as a basis.
Date
8.6.2015
Signature
Holger Asmussen
General Manager
10
Subject to change without notice
General information regarding the
CE marking
HAMEG instruments fulfill the regulations of the
EMC directive. The conformity test made by HA-
MEG is based on the actual generic and product
standards. In cases where different limit values
are applicable, HAMEG applies the strictest
standard. For emission the limits for residential,
commercial and light industry are applied. Re-
garding the immunity (susceptibility) the limits
for industrial environment have been used.
The measuring and data lines of the instrument
have much influence on emission and immunity
and therefore on meeting the acceptance limits.
For different applications the lines and/or cables
used may be different. For measurement opera-
tion the following hints and conditions regarding
emission and immunity should be observed:
1. Data cables
For the connection between instruments resp.
their interfaces and external devices, (computer,
printer etc.) sufficiently screened cables must
be used.
Maximum cable length of data lines must not ex-
ceed 3 m. The manual may specify shorter lengths.
If several interface connectors are provided only
one of them may be used at any time.
Basically interconnections must have a double
screening. For IEEE-bus purposes the double
screened cable HZ72 from HAMEG is suitable.
2. Signal cables
Basically test leads for signal interconnection
between test point and instrument should be
as short as possible. Without instruction in the
manual for a shorter length, signal lines must be
less than 3 meters long.
Signal lines must be screened (coaxial cable -
RG58/U). A proper ground connection is required.
In combination with signal generators double
screened cables (RG223/U, RG214/U) must be used.
3. Influence on measuring instruments.
In the presence of strong high frequency electric
or magnetic fields, even with careful setup of the
measuring equipment an influence can not be
excluded.

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis