Post-use Maintenance
● Immediately after cleaning, disinfecting and drying cycle is complete, the contra angle head must
be removed from the washer-disinfector (within 1 hour) to prevent corrosion�
● After using the washer-disinfector, dry the product completely� Remaining moisture causes internal
corrosion, etc�
● After cleaning and disinfecting the contra angle head, be sure to perform "6 – 3 Lubrication" and
"6 – 4 Sterilization"�
[[For motor handpiece]]
● The motor handpiece cannot be cleaned with washer-disinfector�
[Cleaning]
1
Wipe the debris on the surface of the motor
handpiece with a cloth moistened with water�
[Disinfection]
1
Wipe the surface of the motor handpiece
with disinfectant wipes (Minuten
Wipes:ALPRO)�
CAUTION
CAUTION
2
Under appropriate lighting (≥500lx), inspect
the motor handpiece for contamination�
If any visible contamination remains, repeat
the process from procedure No� 1 until it is
visually clean�
≥ 500lx