Warning: Masks
Always insert the Pediflow
opening provided and do not use any cream in this area
since otherwise the mask may slip from the Pediflow
The masks should not be reprocessed and reused. They
are only intended for single use. Reprocessed masks may
harden and injure the patient.
(4)
Use the headgear to secure the mask and the
Pediflow
straps of the headgear through the loops of the
holder and secure these using the Velcro closures.
In doing so, the headgear should not be attached
too firmly.
Important note: Headgear
Securing the headgear too tightly leads to deformation of
the patient's head.
If the headgear is secured too loosely, adequate fixation
of the Pediflow
(5)
Secure the two lower Velcro straps under the ears
and secure the two upper straps above the ears.
The strap located on the top should be positioned
over the head and forehead. Ensure that the straps
do not go across the eyes. Do not pull the straps
too tightly; they should only lie gently on the skin.
Important note: Securing the mask
(6)
Position the tubing over the baby's head. Two
attachment strips are enclosed for this purpose.
Then secure the inspiration and expiration tubing to
a tube holder and check the position of the
headgear, the Pediflow
Important
expiration tubing
Downloaded from
www.Manualslib.com
Manual
®
completely in the mask
®
to the patient. To do this, thread the
®
cannot be ensured.
Do not pull the bands too tight.
If the leak is too large, check the mask and
®
– pulling the straps more tightly
the Pediflow
does not by itself definitively solve the
problem.
Excessive tightening of the straps can lead to
pressure points on the baby's nose or skin.
®
and the mask once again.
note:
Securing
the
inspiration
Ensure that the weight of the tubing does not
press directly on the Pediflow
the patient's head, but that it can instead be
relieved using tube holders or the pad.
The inspiration and expiration tubing must
slope away from the patient and the
Pediflow® so that any water condensation
does not come into contact with the patient.
medin Medical Innovations GmbH  Adam-Geisler-Str. 1  D-82140 Olching
+49 8142-44846-0  www.medingmbh.com 
®
OP_Pediflow
manuals search engine
®
.
and
®
and thus on
info@medingmbh.com
Ventilation device
Figure 3: System set-up