3. Power Consumption: Smaller than 30mA.
4. Resolution: 1 % for SpO
5. Measurement Accuracy: ± 2 % in stage of 70 % -
100 % SpO
, and meaningless when stage being
2
smaller than 70 %. ± 2 bpm or ± 2 % (select larger)
for Pulse Rate.
6. Measurement Performance in Weak Filling Condi-
tion: SpO
and pulse rate can be shown correctly
2
when pulse-filling ratio is 0.4 %. SpO
pulse rate error is ± 2 bpm or ± 2 % (select larger).
7. Resistance to surrounding light: The deviation
between the value measured in the condition of
man-made light or indoor natural light and that of
darkroom is less than ± 1 %.
8. It is equipped with a function switch. The Oximeter
can be powered off in case no finger is in the
Oximeter within 5 seconds.
9. Optical Sensor
Red light (wavelength is 660 nm, 6.65 mW)
Infrared (wavelength is 880 nm, 6.75 mW)
and 1 bpm for Pulse Rate.
2
error is ± 4 %,
2
38