the SpO
waveform (PLETH) will decrease. In this
2
case, the measurement will be more sensitive to
interference.
2. For those with a substantial amount of staining
dilution drug (such as methylene blue, indigo
green and acid indigo blue), or carbon monoxide
hemoglobin (COHb), or methionine (Me+Hb)
or thiosalicylic hemoglobin, and some with icterus
problem, the SpO
may be inaccurate.
3. The drugs like dopamine, procaine, prilocaine,
lidocaine and butacaine may also be a major
factor blamed for serious error of SpO
4. As the SpO
2
judgement of anemic anoxia and toxic anoxia,
some patients with serious anemia may also report
good SpO
measurement.
2
4 Technical Specifications
1. Display Format: OLED Display;
SpO
Measuring Range: 0 % - 100 %;
2
Pulse Rate Measuring Range: 30 bpm - 250 bpm;
Pulse Wave Display: columniation display and the
waveform display.
2. Power Requirements: 2 x 1.5 V AAA alkaline
battery (or using the rechargeable battery instead),
adaptable range: 2.6 V ~ 3.6 V.
determination by this monitor
2
value serves as a reference value for
37
English
measure.
2