Herunterladen Inhalt Inhalt Diese Seite drucken

Switch Contact; Interrupt Switch; Care And Maintenance; Environmental Note - elector μS-Control 6C Gebrauchsanweisung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Conductivity meter µS-Control 6C
3.5 Switching contact
An external 24 V device can be powered by use of the switching contact. As soon as the
switching point is exceeded this switching contact is de-energized.
The switch contact is suitable to operate a solenoid valve (normally closed). Suitable solenoid
valves are available from us.

3.6 Interrupt switch

An interrupt switch can be found on the left side of the housing. By pressing this button you
interrupt the function of the switching contact, provided that the switching point is exceeded. The
switch contact gets energized when the interrupt switch is pressed.
If a solenoid valve is connected to the switch contact, it can be opened by pressing the interrupt
switch.

3.7 Care and maintenance

Clean the housing only with a damp cloth.
Do not use cleaning agents that contain solvents.
The device is maintenance and calibration free.
Conductivity sensors that are used in deionized water are maintenance-free.
Conductivity sensors, that are used in normal water or in heating water, can become dirty, which
means that inaccurate measured values are possible. Clean such sensors at regular intervals with
a cleaner suitable for stainless steel and plastics.

3.8 Environmental note

If the unit is not repairable defective, pay attention to proper disposal separately from household
waste.
Inquire at the city or municipal administration about the options for the environmentally friendly
and appropriate disposal of old electronic devices.
www.elector-water-treatment.com
Page 14

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis