Preventive maintenance

The following components must be maintained:
Sealings for inter-
Every 2 years
mediate plate
CLIC adapter
Maintenance-free, but must be replaced every
4 years
Maintenance must be performed by experts, e.g., DrägerService.
CAUTION
The CLIC adapter is subject to wear due to use and
reprocessing.
Replace the CLIC adapter after 4 years at the latest or when
there are signs of damage and wear.
Disposal
CAUTION
The spent disposable CO
absorber cannot be prepared for
2
reuse.
At the end of life, dispose of the product in accordance with
applicable regulations.
It is recommended that the possibility of disposing of soda lime
in household waste be checked with local waste disposal
authorities.
In Europe, the applicable waste material code for soda lime is
EWC 160303/180106.

Technical data

Ambient conditions
During operation
Temperature
10 °C to 40 °C (50 °F to 104 °F)
Ambient pressure
500 to 1200 hPa (7.3 psi to 17.4 psi)
Relative humidity
0 to 95 %
During storage and transport
Temperature
–20 °C to +50 °C (-4 °F to 122 °F)
CLIC adapter
–20 °C bis +55 °C (–4 °F bis 131 °F)
and intermediate
plate
Ambient pressure
500 to 1200 hPa (7.3 psi to 17.4 psi)
Relative humidity
0 to 95 %
Resistance
Reduced by approx. 1 mbar in the bypass mode.
Depending on the breathing system of the re-
spective Dräger anesthesia workstation.
Compliance
Normal mode: approx. 1.3 mL/mbar.
Bypass function: approx. 0.4 mL/mbar.
Leakage
<5 mL/min.
Classification
Class IIa
In accordance with Directive 93/42/EEC Annex
IX
UMDNS code
17-509
Universal Medical Device Nomenclature System
– Nomenclature for medical devices
The medical device meets the requirements of the ISO 80601-2-
13 standard.
Instructions for use CLIC Absorber 800+ / Infinity ID CLIC Absorber 800+ / CLIC Adapter

List of accessories

CLIC Absorber 800+
Infinity ID CLIC Absorber 800+
Intermediate plate
Cleaning disk
CLIC adapter
Sealing
Sealing set for intermediate plate
Safety data sheet
The Drägersorb 800 Plus safety data sheet is available on
request.
English
MX 00 004
MX 50 004
MX 00 013
MX 50 081
MX 50 090
MX 50 072
M 35 380
13

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Infinity id clic absorber 800+

Inhaltsverzeichnis