WARNING!
These short instructions are intended for commissioning
the Stratos Eco 2405 pH transmitters. They are intended for
experienced users and do not replace the instruction manual
(on included CD-ROM or at www.knick.de).
The shipment includes a special document with safety
information.
The Stratos Eco 2405 pH is used for pH/mV, ORP, and temperature
measurement in industry, environment, food processing, and sewage
treatment.
The sturdy molded enclosure can be fi xed into a control panel or
mounted on a wall or at a post.
The protective hood provides additional protection against direct
weather exposure and mechanical damage.
The device has been designed for application with commercially
available sensors with a nominal zero point at pH 7. It provides
two current outputs (for transmission of measured value and tem-
perature, for example), two contacts, and a universal power supply
24 ... 230 V AC/DC, AC: 45 ... 65 Hz.
Terminals:
Screw terminal, suitable for single wires / fl exible leads up to 2.5 mm
(AWG 14).
Recommended torque for the terminal screws: 0.5...0.6 Nm.
Caution!
Additional safety precautions have to be taken for operation in
hazardous locations according to CSA (CLI, DIV2, GPA,B,C,D T4,
Ex nA IIC T4) (see
special document)!