Herunterladen Inhalt Inhalt Diese Seite drucken

Sl Analytics TitroLine 5000 Gebrauchsanleitung Seite 118

Titrator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

118
The blind value which was possibly titrated in advance, is always taken into consideration
automatically.
Fig. 80
Example: The blank value of a chloride titration is defined with the support of an extra method. The
result in ml is thereby automatically written into global memory M01 by using the name "Blanc value
Chloride". The blank value is then automatically deducted from the titrant consumption within the
chloride
method.
Here in our example it is 0.035 ml:
Fig. 81
4.6.3.6 Statistics
The mean value and relative standard deviation can be automatically calculated and documented by
using the statistics.
Fig. 82