Maintenance / Cleaning
To ensure longevity of the chair and safety for the user, the following should be included in
the daily inspection:
•
Inspect and clean the wheels of dust and dirt
•
Inspect and clean the wheel brakes of dust and dirt
•
Check that all screws and nuts are properly tightened
•
Check that the snaplock clips are securely fastened
After each use, or according to facility protocol, the Combi chair may be cleaned/disin-
fected with non-abrasive, non-corrosive solutions. The PUR seat and backrest can be
detached for cleaning, as can the bucket / bedpan holder (NOTE: the rear section of the
bucket / bedpan holder must be lifted upwards before removing the two rods from the
holes). The entire Combi chair may also be washed / autoclaved at 85 degrees Celsius for
3 minutes. Avoid harsh solvents and chlorine. If alcohol is used to clean the black polyure-
thane, discolouration may occur. Ensure all parts are re-attached correctly prior to use.
Reuse
Follow the above procedures. Defective or worn parts must be replaced.
WARNINGS
Max. load Combi Standard & Low: 150 kg / 330 lbs,
•
Combi Wide: 200 kg / 440 lbs.
•
Check the condition of the snaplocks frequently, and regularly
check that the snaplocks are inserted correctly into the frame.
•
Lock wheels when chair is stationary and also during transfers to and from the
chair.
•
The Combi chair is not designed to be used as seating for extended-periods.
•
The Combi chair should be pulled backwards over thresholds. For optimal
caregiver ergonomics, the push handles should be at elbow height.
•
Always use the COMBI chair with armrests in place (folded down). Armrests (and
footrests, if present) should only be flipped up during transfer to and from the
chair.
•
If the castors are adjusted in height, it is important to mount them again correctly
(see "Assembling the Casters" above). Ensure that all 4 casters are adjusted to
the same height.
•
Keep fingers away from the armrest joints to avoid pinching during raising and
lowering.
•
Use the chair only on a flat, even surface.
•
Insert or remove the bedpan/bucket when the chair is not in use.
•
Do not store the chair in direct sunlight as the black surface may become hot and
uncomfortable for the user.
•
To avoid injury, do not insert fingers or toes into the drainage holes in the backrest
or footrests.