Troubleshooting
Control of the operation panel on the Camera Unit is impossible
Possible Cause
The Remote Controller has been con-
nected.
The priority of the control is given to the
Remote Controller when the power is
turned on.
Damage in the Camera Cable.
Control of the operation on the Remote Controller unit is impossible
Possible Cause
The priority of the control is given to the
Camera Unit.
Damage in the Remote Controller Cable.
Page 8-30
Measures
Push the REMOTE switch
to change the control pri-
ority to the Camera Unit.
We recommend the Re-
mote Controller be dis-
connected when not in
use.
Contact your local
Distributor.
Measures
Push the REMOTE switch
to change the control pri-
ority to the Remote Con-
troller.
Contact your local
distributor.