MOTOR BRUSHES MAINTENANCE
With your Eolo Plus aspiration unit, you will receive two spare bruses for the motor. The unit has a 16000 rpm
motor, so brushes must be replaced periodically. After working 1000 h, the electronic controls will show the
message 'br', meaning that it is the time to replace the brushes. Then, proceed as following:
1. Unplug the unit.
2. Release the 6 bolt of the inner lid and remove it. You will get access to the motor.
3. Remove the 4 bolts holding the brushes (10).
10
10
11
4. Remove the brushes and disconnect the cables (11).
5. Disasemble metal parts from plactic parts (12).
12
13
6. Replace the new brushes into the plastic parts (13).
7. Connect the cables to the new brushes (14).
14
15
8. Use a screwdriver to pull the grey part of the brushes and make them go out (15).
9. Replace the new brushes on the motor and fix them with the 4 bolts (10).
10. Close the inner lid and check the correct working of the unit.
11. In order to reset the brush timer, press simultaneously keys [AUTO/MAN] and [–] for 5 seconds. Notice
how the display blinks and 2 seconds later the timer resets and the 'br' message will dissapear.
NOTE: When the message 'br' appears, you will have to put the unit off and on in order to run again. The
message will keep on to remind that brushes must be replaced as soon as possible.