Herunterladen Diese Seite drucken

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27

Pump Fixing

WARNING!
If a rotor failure occurs, the connection of the pump to the system could be
subjected to a significant torque. If the connection is not sufficient to
withstand that torque, the pump could detach from the system or the motor
housing could detach from the pump envelope. In this case metal fragments
could be projected from the pump or system, which could cause serious
injury or death and/or damage to surrounding equipment.
Fix the TV 801 Navigator Bolted Flange in a stable position, mounting
the inlet flange of the turbopump to the system counter-flange, with a
connection capable of withstanding a torque of 8900 Nm around its
axis. For example, this Bolted Flange can be fixed using high strength
steel screws (class higher than 8.8).
The following table shows the necessary number of screws and the
relevant fixing torque.
Tab. 1
FLANGE
160 Bolted
The turbopump with ConFlat inlet flange must be fixed to the
vacuum chamber by means of the appropriate Agilent hardware. See
the appendix "Technical Information" for a detailed description.
The TV 801 Navigator Bolted Flange can be installed in any position.
The TV 801 Navigator Bolted Flange cannot be fixed by means of its base.
NOTE
TV 801 Navigator Bolted Flange User Manual / 87-900-980-01
N.
FIXING TORQUE
8
5 Nm
(D)
Instructions for Use

Pump Fixing

201/248
16

Werbung

loading