Instructions for Use
Installation
The TV 301SF Navigator can be installed in any position. Fix the TV
301SF Navigator in a stable position connecting the inlet flange of the
turbopump to a fixed counter-flange capable of withstanding a torque
of 1000 Nm around its axis.
The turbopump with ISO inlet flange must be fixed to the vacuum
chamber by means of clamps or claws. The following table shows, for
each flange and fixing device, the necessary number of clamps or
claws and the relevant fixing torque.
Tab. 1
FLANGE
ISO 100
See the appendix "Technical Information" for a detailed description.
The TV 301SF Navigator cannot be fixed by means of its base.
NOTE
The TV 301SF Navigator belongs to the second installa-tion (or overvoltage)
CAUTION!
category as per directive EN 61010-1. Connect the device to a mains line that
satisfy the above category. The TV 301SF Navigator has Input/Output and serial
communication connectors that must be connected to external circuits in such a
way that no electrical part is accessible. Be sure that the insulation of the device
connected to the TV 301SF Navigator is adequate even in the case of single fault
as per directive EN 61010-1.
For installation of optional accessories, see "Technical Information".
230/300
FIXING DEVICE
M10 clamps
M8 claws
TV 301SF Navigator User Manual / 87-900-957-01 (H)
N.
FIXING TORQUE
4
22 Nm
8
11 Nm