4.2.
Camera head description
[12]
[13]
[14]
5. Installation
No special training is required to install this medical device. Please refer to the instructions in this section of the
user manual.
Once all iCare components have been unpacked from their original packaging, please perform the following ac-
tions:
- Place the control unit on a level and stable surface. If you place it in a compartment, make sure that it is
sufficiently ventilated (at least 15cm around the unit).
- Connect the power cord to the power outlet [2] on the rear of the unit.
- Connect the other end of the power cord to an electrical outlet.
- Connect the HDMI cable to the corresponding output [3] on the rear of the unit.
- Connect the second end of the HDMI cable to the corresponding input on the monitor.
- Place the two WiFi keys on the two USB outputs [5] on the rear panel.
- Place the USB key on the USB output [4] on the rear panel.
- Set the power supply button on the rear panel of the unit to position « I ». The unit enters in a booting
sequence: the indicator light [9] on the front panel flashes rapidly. This booting sequence ends after approxima-
tely 1 minute 30. Then, the unit goes into standby mode: the front panel indicator light flashes slowly.
- Power on the monitor. A colour bar should appear on the screen.
- Plug the camera head connector to the front of the control unit [10]. A red coding is present on the
connector of the camera head and above the corresponding socket of the control unit. Align these two coded pins
to connect the camera head. A « click » lock must be heard.
- Insert the light cable into the hole provided for this purpose [8]. A « click » lock must be heard. Connect
the other end of the light cable to your endoscope.
- Place an endoscope in front of the camera lens.
- Press the stand-by button [11] on the right side of the control unit to start the device: the indicator light
on the front panel becomes fixed. Another press will put it back on standby.
It is then possible to check the correct operation of the camera and especially to ensure that the settings of the
monitor give full satisfaction thanks to the colour bar which appears when disconnecting the sensor.
D900 700 053 B
Instructions for use: iCare imaging station
[12]
: This button allows to launch a white balance by long press.
[13]
: This button allows to switch from a minimum value of light power to a maxi-
mum value by short press, and vice versa. These values are 75% and 100% of the light
output respectively.
[14]
: This button allows:
- by short press, to launch an image capture,
- by long press, to start or stop a video recording.
Rev 09/2018
delmont
imaging
8