Safety test
Severity Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Note 1: At 80 MHz and 800 MHz, the highest frequency band should be used.
Note 2: These recommendations may not be applicable in all situations. The propagation of electromagnetic
waves is altered by absorption and reflection from structures, objects and people. For transmitters whose maxi-
mum output power is not listed in the table above, the recommended separation distance d, in metres (m) can be
established using the equation applicable to the transmitter frequency, where P is the maximum output power of
the transmitter in Watts (W) assigned by the transmitter manufacturer.
11.3. Recommended distances between portable and mobile RF communication systems
for this product
The « CMOS camera + LED source » reference equipment has been designed for use in the electromagnetic
environment in which the emitted RF interference is controlled. The user of this equipment can help avoid elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communication
systems (transmitters) and this equipment, as recommended below, as a function of the maximum output power
of the communication system.
Maximum assigned
transmitter output power
in W
0.01
0.1
1
10
100
D900 700 053 B
Instructions for use: iCare imaging station
IEC 60601
Compliance
Level
3V
3 V/m
Separation distance as a function of transmitter frequency (m)
150 kHz to 80 MHz
d = 1,16
P
0.116
0.366
1.16
3.66
11.6
Rev 09/2018
Electromagnetic environment: Guidance
Portable and mobile RF communication devices should
not be used at a distance, including cables, from this
product that is less than the recommended distance,
calculated by applying the formula that corresponds to
the transmitter frequency.
d = 1,16
P
d = 1,16
P 80 MHz to 800 MHz
d = 2,33
P 800 MHz to 2,5 GHz
Where P is the maximum output power of the trans-
mitter, in Watts (W), assigned by its manufacturer and
(d) is the recommended separation distance in metres
(m).
Field strength levels emitted by fixed RF transmitters -
which must be established by in situ electromagnetic
measurement - must be below the compliance level in
each frequency band. Interference may occur with de-
vices marked with the following symbol:
80 MHz to 800 MHz
d = 1,16
P
0.116
0.366
1.16
3.66
11.6
delmont
imaging
800 MHz to 2.5 GHz
d = 2,33
P
0.233
0.736
2.33
7.36
23.3
14