POSSIBLE PROBLEMS AND HOW TO SOLVE THEM

If the device does not switch on:
1.
make sure the plug is firmly fitted to the wall socket;
2.
the thermal protector might have been released. The thermal protector may be released
for the following reasons:
- the device has been in operation for a too long time;
- the device was used near heaters or in hot environments;
- the air vent openings (pic. A/7) are obstructed and do not allow for the motor ventila-
tion.
In this case, set the ON/OFF switch to "O" and let it rest for at least 60 minutes.
If the device does not nebulize, or only very weakly, check the following
points:
1.
make sure that the ends of the air tube are fitted tightly onto the main unit and onto the
nebulizer;
2.
check whether the nebulizer is empty from medication or full of an excessive medication
quantity (MAX 15 ml);
3.
check whether the nebulizer nozzle is obstructed;
4.
check whether the air conduction cone is inserted (pic. E).
Should the device not start working properly again, address immediately to
an authorized technical service centre or to MEDEL CUSTOMER SERVICE.

MAINTENANCE AND REPAIRS

In case of failure, address to qualified personnel authorized by Medel S.p.A.. Never open
the device. There are no user-serviceable parts within. The compressor requires no lubrica-
tion or maintenance.
man PRO EN 11-18.indd 15
man PRO EN 11-18.indd 15
EN
15
3-06-2008 13:15:33
3-06-2008 13:15:33