inhalation valve: remove it (pic. B/16) and connect the nosepiece or the masks directly
onto the nebulizer (pic. I).
10. Plug the device to the wall socket, making sure that the electrical rating shown on the
rating plate on the bottom of the unit corresponds to your main voltage.
11. To start the treatment, switch the device on by pressing the switch (pic. A/2) in the front
of the device onto the "I" position.
12. Inhale the aerosol solution using the prescribed accessory.
13. When the treatment has been completed, switch the unit off by setting the switch position
"O" and disconnect the plug from the wall socket.
14. Wash the nebulizer and accessories as described in the "CLEANING AND DISINFECTION
OF THE ACCESSORIES" section.
15. Put the nebulizer, accessories and power cord inside the carrying case (pic. A/15).

CLEANING AND MAINTENANCE

CLEANING OF THE MAIN UNIT
Use a clean dry cloth and mild detergents for cleaning the unit.
ATTENTION: during cleaning operations, make sure liquids do not get
inside the device.
CLEANING AND DISINFECTION OF THE ACCESSORIES
After each treatment:
The nebulizer MEDELJETPRO, the mouthpiece and the nosepiece must be disinfected by boi-
ling in water (for max. 10 minutes) or in autoclave (max 121°C, see autoclave instructions for
duration). We suggest you disinfect the masks using cold disinfecting liquids (solutions with
hypochlorite). Dry the nebulizer using a clean cloth; after connect it to the air outlet (pic. A/4),
switch the device on and let it work freely for 5-10 minutes.
Do not boil nor autoclave the air tube and the masks.
Following carefully the cleaning and disinfecting instructions is very
important for the product performance and for a successful therapy.
REPLACEMENT OF THE NEBULIZER
The nebulizer must be replaced after a long period of inactivity, or when it is obstructed by
dry medication, dust, etc.
Your nebulizer should last from 6 months to 1 year with general use.
Use original nebulizers only.
REPLACEMENT OF THE FILTER
The MEDEL PRO Porex filter must be replaced approximately after 500 hours of use or each
year; control the filter periodically (10-12 aerosol treatments ); when it turns grey or brown or
wet, replace it; open the filter holder (pic. A/5) and replace the filter with a new one. Don't
try to clean the filter and to reuse.
Use original filters only.
Do not use the unit without filter.