Herunterladen Inhalt Inhalt Diese Seite drucken

Installation - Sunoptic Surgical LED-3000 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für LED-3000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 29

5. INSTALLATION

5.1 SETTING UP THE LED LIGHT SOURCE
Place the LED light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or flammable
gases.
DO NOT block the venting grids of the light source or impede exhaust air
flow. Users must ensure that ventilation air temperatures are below the
maximum rating for the light source and are not influenced by warm
exhaust from other sources.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light
source.
Use only cords provided with the light source.
CAUTION
Plug the AC power cord into a wall outlet using the three-prong plug supplied
with the unit.
To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.Make sure that the power cord is connected to
the main power with three-prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
5.2 CONNECTING THE FIBER OPTIC CABLE
Connect the light cable to the instrument and then plug the light cable end fitting into
the turret port on the front panel.
NOTE
The turret has a cable in place sensing feature that will not allow the LED
light to turn on without a fiberoptic cable inserted into the turret port.
LIT195 SUNOPTIC SURGICAL
(English)
®
Rev. C
Date of Revision:9/15/15
Page 8 of 78

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis