Herunterladen Inhalt Inhalt Diese Seite drucken

Hach FL1500 Basis-Benutzerhandbuch Seite 18

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
Signal
US9000
2 V –
Blue
1 SRQ
Installation for Flo-Dar or Flo-Tote sensors
Install the air reference tube and external desiccant to make sure that the pressure transducer in the
sensor operates correctly. The desiccant cartridge prevents damage from moisture and debris.
Moisture and debris can decrease the accuracy of the Flo-Dar and Flo-Tote sensors.
If the instrument is installed in a high-moisture environment, install a desiccant cartridge.
1. Remove the plug from one of the air reference ports. Refer to
Note: Do not remove the plugs from the air reference ports that are not used.
2. Push the air reference tube from the sensor into the air reference port.
Note: If the sensor cable is pulled through a conduit, protect the wiring and air tube from damage and
contamination.
3. Install an external desiccant cartridge on the air inlet port. Refer to
on page 30.
Figure 9 Air reference tube and desiccant installation
1 Air inlet
2 Desiccant cartridge
18 English
Table 4 Sensor wiring (continued)
2
pHD
BL9000
AV9000S
Black
Green
Blue
3 Plugs—remove only for air reference tubes
4 Air reference tube from sensor
3
Flo-Dar
SVS
Flo-Tote 3
Green
Green
Figure
9.
Figure 9
and
Junction box
for US9000
Black
Accessories

Werbung

Kapitel

Inhaltsverzeichnis
loading