Routine maintenance

Routine maintenance must be entrusted to specially instructed
surgery staff.
• We recommend to pay special attention to all danger signals
and to use goggles, gloves and disposable overalls for per-
sonal protection.
Every evening
• at the end of every working day aspirate a solution of Puli-Jet
plus disinfectant using Pulse Cleaner;
• disconnect the machine from the mains before any interven-
tion.
Daily
• place the Antifoam Tablets on the dental unit fi lters.
Periodically
• keep the aspirator fi lter clean.
Occasionally
• make sure that the aspirator ventilation is not obstructed;
• keep the plant room free from anything not related to the
machines, especially from fl ammable material. Make sure that
there is no possibility for the formation of corrosive, fl ammable
and explosive mixtures.