Approvals

2.2

Intended Use

The Dräger VVP 1000 (hereafter referred to as "display") is a
display for the graphical visualisation of up to 100 channels of
a VarioGard system.
2.3

Limitations on use

The display is not designed or authorised for use in areas
where flammable or explosive gas mixtures could develop.
3
Approvals
Safety
IEC 61010-1, EN 61010-1: Safety requirements for electrical
components, Part 1.
IP66 degree of protection classification (front only), IEC 529
Type 4X degree of protection classification (front only), UL50
Electromagnetic compatibility
Emissions and immunity for EN 61326: Electrical equipment
for measurement, control and laboratory use.
Immunity to industrial locations:
RF interference
EN 61000-4-6 Criterion B
Surge voltage
EN 61000-4-5 Criterion A
Fast transients (burst) EN 61000-4-4 Criterion A
Power
frequency
EN 61000-4-3 Criterion A
magnetic fields
Electrostatic
EN 61000-4-2 Criterion A
discharge
Emissions:
EN 55011
Class A
20
3 V/rms
1 kV L-L,
2 kV L&N-E power
2 kV power
1 kV signal
10 V/m
8 kV air discharge
4 kV contact discharge
FCC conditions
This device has been tested and complies with Part 15,
Class B of the FCC conditions. Its operation is subject
to the following conditions:
(1) This device must not cause harmful interference.
(2) This device must accept any received interference;
this includes interference caused by unintentional
operation.