CONTROL SOFTWARE

Easy to install and use, the control software will run under Windows 98, 95 or 3.1. And enable the user to control one or more EasyLogs and operate
them as a complete system. Supplied on a 3½" disk with a manual and serial link.
LINK FUNCTIONS
Lk2 & Lk3: Fit these jumpers as shown in the table below.
Measurement Range Jumper
200mV
2V
20V
4-20mA
Temperature (Low Range)
Temperature (High Range)
PIN FUNCTIONS
Vin- Voltage measurement Input. (see Applications below)
Vin+ Voltage measurement Input. (see Applications below)
I
out Current loop connection. (see Applications below)
I
in Current loop connection. (see Applications below)
CTS RS232 Communications connection. (Connect to Pin 7 of 9-way D-Type)
DSR RS232 Communications connection. (Connect to Pin 4 of 9-way D-Type)
TX RS232 Communications connection. (Connect to Pin 2 of 9-way D-Type)
RX RS232 Communications connection. (Connect to Pin 3 of 9-way D-Type)
GND RS232 Communications connection. (Connect to Pin 5 of 9-way D-Type)
BATTERY REPLACEMENT
Only use 3.6V C-size Lithium Thionyl Chloride batteries. (e.g. Sonnenschein part number SL 770/S). Check with your supplier that the battery you
are ordering is 'press fit' and is not fitted with solder tags or pins. Take care to connect correctly.
WARNING: Handle lithium batteries carefully - observe warnings on battery casing.
IP67 / NEMA 4X RATING
The connectors featured on EL-IP67-12 exceed enclosure rating IP17 when not mated or covered and IP67 when mated or covered (IEC 529). The
abrasion resistant thermoplastic boot provides strain relief and accepts cable diameter between 0.195" and 0.265". The paint coating on the metal
housing of the EL-IP67-12 must be unscratched in order to achieve a NEMA 4X rating.
APPLICATION
NOTE -
circuit to convert the parameter to be measured into a linear voltage or current and apply this signal to a suitably scaled EL-IP67-12 module.
Measurement signals must always be isolated from the communications signals.
CE
Complies with EN 61326-1 (1997) Electrical Equipment for Measurement, Control and Laboratory Use.
EL-IP67-12 is an easy to use data logging module capable of measuring, recording and displaying temperature (Pt
100), voltage and current. With a 12 bit A/D converter and a memory for 8000 readings, EL-IP67-12 can operate
as a 'stand alone' logger or be permanently connected to a system. The data logger is housed in an IP67 / NEMA 4X
rated enclosure to protect it from ingress of dirt and moisture. The PC software operates under Windows and does
not require specialist skill to operate. Communication between PC and data logger can take place via the RS 232
or InfraRed ports. The EL-WIN control software allows the data to be output in text format for easy integration into
most popular spreadsheets. Graphical output is also possible under EL-WIN. Consult the EasyLog software
manual for further details.
IP67 / NEMA 4X Protection
InfraRed Communications
Easy to Use
Stock Number - EL-WIN
Link Setting
Lk2 open, Lk3 open
Lk2 closed, Lk3 open
Lk2 closed, Lk3 open
Lk2 open, Lk3 closed
Lk2 open, Lk3 open
Lk2 open, Lk3 open
Dispose of in accordance with local regulations.
It is possible to measure parameters other than those outlined below. Use an appropriate sensor and conditioning
High Resolution Read-out
Non-volatile Data Storage
Battery Powered
2
Sensor Connection
Vin+
Vin-
Iout
Iin
Female Socket - Front View
Communications Connection
Male Socket - Front View
CTS
DSR
Not
Used
RX
GND
TX