Important safeguards and notices for use
in the medical environments
1. All devices connected to the unit must be certified or
compliant according to IEC 60601-1, IEC 60950-1,
and IEC 60065 standards and other IEC/ISO standards
applicable to the devices.
2. Furthermore, the system as a whole must comply with
IEC 60601-1 standards. All peripheral devices
connected to the signal input/output sections of the unit
constitute the medical-use system, and therefore, the
user is responsible for ensuring that the system as a
whole complies with IEC 60601-1 standards. If in
doubt, consult qualified Sony service personnel.
3. Connecting the unit to other devices may increase the
leakage current.

Important EMC notices for use in medical environments

• The HMM-3000MT needs special precautions regarding EMC and needs to be installed and put into service according
to the EMC information provided in the instructions for use.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the HMM-3000MT.
WARNING
The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased immunity of the HMM-3000MT.
Type of cable
HMO-CA50M : Head Mounted Display Cable
Guidance and manufacturer's declaration – electromagnetic emissions
The HMM-3000MT is intended for use in the electromagnetic environment specified below. The customer or
the user of the HMM-3000MT should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
List of cables used for EMC test
Compliance
Group 1
Class B
Not applicable
Not applicable
4. For all peripheral devices connected to the unit that
operate on commercial power supplies and do not
comply with IEC 60601-1 standards, incorporate an
isolation transformer that complies with IEC 60601-1
standards and connect to the commercial power supply
via the transformer.
5. The unit generates, uses, and may radiate radio
frequency energy. If it is not installed and used in
accordance with the instruction manual, it may cause
interference on other devices. If the unit causes
interference (which can be determined by
disconnecting the power cord from the unit), try the
following.
– Relocate the unit with respect to the affected devices.
– Connect the unit and the affected devices to different
branch circuits.
For more information, consult qualified Sony service
personnel.
(Applicable standard: IEC 60601-1-2)
Specifications
5 m, Shield cable
Electromagnetic environment – guidance
The HMM-3000MT uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The HMM-3000MT is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.