ACTIVE AND PASSIVE TEMPERATURE AND HUMIDITY TRANSMITTER
DO 9861T humidity and temperature transmitters convert the humidity and
temperature measured by two sensors into a 4÷20 mA signal.
An LCD indicator allows the value of the process signal and the various
parameters to be displayed.
The accurate design and choice of components make the instruments precise
and reliable for a long working life. The R.H. sensor is temperature compensated.
The instruments have two sensors, a capacitive sensor for measuring relative
humidity and a Platinum resistive sensor (Pt100, 100 Ω at 100°C) for measuring
temperature.
The transmitters measure the relative humidity (% RH), the sensor temperature
(TC), and calculate the wet bulb temperature (WT), the dew point temperature
(DP), the absolute humidity (gr/m3) and the mixing ratio (gr/kg - grams of water
per kg of dry air). The temperatures are expressed in degrees Celsius.
The measurements refer to an atmospheric pressure of 1013.25 mbar. Once
the physical quantity to be checked has been selected, the output current will be
proportional and the status of the relays will depend on that variable.
Technical characteristics DO 9861T
Capacity
Transducer energizing
Cable length
Accuracy
Working range of sensor
During use, check that the sensor is compatible with the atmosphere used
2/4 wires Pt100
Transducer energizing
Cable length
Accuracy
4,00...20,00 mA
Accuracy
Load resistance
A and B
Passive
Active
Without filter
With filter
Front panel
External dimensions
Protection class
* time required to reach 63% of the final variation.

Key functions

PRG Programming of the parameters is activated by pressing the PRG key
along with the ▲ and ▼ keys. The message P1 appears on the display
to indicate that parameter P1 is being programmed. When the PRG key
is pressed in sequence the messages P2, P3, etc. appear as well as the
corresponding parameters. After P13 the instrument returns to normal
function.
300 pF
10 kHz
<10 m multiple screen (about 2 nF)
additional error about 0.003 pF/pF
Instrument: 0.1% of reading ±1 digit ±0.01%/°C
Sensor: ±2.5% (5...90%), +3/-2.5% (90...98%)
Relative humidity: 5...98% @ 23°C
Dew point temperature (DP°C):
- 60..+100 DP°C (-20 DP°C @ 23°C)
Temperature range: -40..+150°C
-50...+199.9°C
0.5 mA dc
<10 m not screened
<50 m screened (about 5 nF)
0.2°C ±0.1% of reading ±0.01°C/°C
Dry bulb temperature -50.0...+199.9°C
Wet bulb temperature -50.0...+199.9°C
Relative humidity 0.0...+100.0% RH
Absolute humidity 0.0...+1999 gr/m3
Mixing ratio 0.0...+1999 gr/Kg of air
±2.5% RH
0.5% of reading ±0.02 mA
= Vdc-9
R
R
= 636 Ω @ Vdc = 24 Vdc
Lmax
Lmax
0,022
Bistable exchange contacts 3 A/230 V ac potential
free contact
4÷20 mA, 10...35 V dc, see figure 2
24 or 230 V ac -15/+10%, 1 VA, 48...62 Hz, see
figure 1
6 seconds
3 minutes
122x120 mm
122x120x56 mm
IP64
SET Key for setting the relay intervention threshold. The ON or OFF symbol
appears on the display to indicate that the instrument is showing the
switching on or the switching off threshold of relay A or relay B.
TD/TW When this key is pressed the dry bulb temperature TD or the wet bulb
temperature TW is shown at the top of the display
UNIT - When this key is pressed the dew point temperature (DP), the relative
humidity (% RH), the absolute humidity (gr/m3 - grams of water per cubic
metre of dry air) or the mixing ratio (gr/kg, grams of water per kg of dry air)
EXAMPLES OF CONNECTION OF THE DO 9861T
��
� ��
��
��
����� � �����
��
��
��
��
����
����� ������� ������� ��� �� ��� �� �� ����� �� �� �������� ����������
Fig. 1: Connection of the humidity and temperature sensor, screened cable, Pt100 4 wires +
screen, active transmitter with alternating power supply, double analog output 4÷20 mA
��
� ��
��
��
����� � �����
��
��
��
��
����
Fig. 2: Connection of the humidity and temperature sensor, screened cable, Pt100 2 wires,
passive transmitter, fed with 3 wires by a regulator/indicator, double analog output 4÷20
mA.
��
��
��
��
��
��
��� �
���
��
��
��
��
����� � ������� ������
����� ��
����� ���
�� ��
�� ��
���
��
��� �� � ��
���� ��
���