EMC Declaration
Recommended separation distances between portable and mobile RF
communications equipment and the blood pressure monitor.
The "blood pressure monitor" is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer
or the user of the blood pressure monitor can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the "blood pressure monitor"
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
0,01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
eparatlon distance according to frequency of transmitter
150 kHz to 80 MHZ 80 MH to 800 MHZ 800 MHz to 2,5 Ghz
0.12
0.38
1.2
3.8
12
m
0.12
0.38
1.2
3.8
12
24
0.23
0.73
2.3
7.3
23