Type of error
Battery symbol appears
on the display
Er 0
Er 1
Er 2
Er 3
Er 4
Er 5
Er 6
Er 7
The entire display
is flashing
Er 9
Er A
Er B
The appliance does
not respond when
the keys are pressed
20

TROUBLESHOOTING

Cause
Run-down batteries
The user has moved
during the measurement
Error in the measurement
of the systolic pressure
Error in the measurement
of the diastolic pressure
Inflation blocked because
the wristband is too tight
Air leak or the wristband is
too loose
Inflation above 300 mmHg
Inflation above 15 mmHg
after 3 minutes
Error in accessing the
memory system
Error in the control of
the appliance parameters
Error in the self-diagnosis
of the MCU
Capacity error
Error when saving in the
memory system
Incorrect operation
Checks
Replace the batteries
Do not move, and repeat
the measurement
Repeat the measurement
Repeat the measurement
Reposition the wristband
correctly and check the connection
Reposition the wristband
correctly and check the connection
Press the START key
Press the START key
Contact the retailer
Contact the retailer
Contact the retailer
Contact the retailer
Contact the retailer
Remove the batteries for
5 minutes and then reinstall
them