Herunterladen Inhalt Inhalt Diese Seite drucken

OHAUS SF40A Bedienungsanleitung Seite 19

Vorschau ausblenden Andere Handbücher für SF40A:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
EN-16
Problem / Message
Not Executable
--------
Date/Time Not Set
----
Memory Error
---------
Memory full
---------
Load/Unload Weight
--
Unit change not
allowed!
Out of range
---------
Possible cause
 Instrument cannot
------
execute printer
instruction.
 Printer error
---
messages.
---------
 Statistics or
----------
Totaling Memory is
full.
 No sample or the
---
same sample is on
the weighing pan.
 Changed unit
during Statistics or
Totaling
application.
 Sample weight is
---------
out of range
SF40A Printer
Correction
 Check the balance.
 Check the instrument.
 Reset the printer.
 Set date/time
 Back-up battery
depleted
Please contact OHAUS
service.
 Start a new Statistics or
Totaling application.
 Put a new sample or
remove it and put it
again on the pan.
 Reset the balance.
 Change the unit back to
the first one.
 Sample weight must be
in the range of 70% -
130% of the current
average value.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis